4.0 Article

An autosampling differential scanning calorimeter instrument for studying molecular interactions

期刊

出版社

MARY ANN LIEBERT INC PUBL
DOI: 10.1089/154065802761001338

关键词

-

向作者/读者索取更多资源

A new ultrasensitive differential scanning calorimeter (DSC) instrument is described, which utilizes autosampling for continuous operation. High scanning rates to 250 deg/h with rapid cooling and equilibration between scans facilitates higher sample throughput up to 50 samples during each 24 h of unattended operation. The instrument is suited for those pharmaceutical applications where higher throughput is important, such as screening drug candidates for binding constant or screening solution conditions for stability of liquid protein formulations. Results are presented on the binding of five different anionic inhibitors to ribonuclease A, which included cytidine 2'-monophosphate (2'CMP), 3'CMP, uridine 3'-monophosphate, pyrophosphate, and phosphate. Binding constants K-B (or dissociation constants K-d) are obtained from the shift in the transition temperature T-M for ribonuclease thermal unfolding in the presence of ligand relative to the transition temperature in the absence of ligand. Measured binding constants ranged from 155 M-1 (K-d = 6.45 mM) for the weak-binding phosphate anion to 13,100 M-1 (K-d = 76.3 muM) for the strongest binding ligand, 2'CMP. The DSC method for measuring binding constants can also be extended to ultratight interactions involving either ligand - protein or protein - protein binding.

作者