Accuracy of displayed values of tidal volume in the pediatric intensive care unit

期刊

CRITICAL CARE MEDICINE
卷 30, 期 11, 页码 2566-2574

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/00003246-200211000-00027

关键词

tidal volume; intensive care units; pediatric; ventilators; mechanical; evaluation studies; infant; child; intubation; intratracheal

向作者/读者索取更多资源

Objectives: To assess the accuracy of the expired tidal volumes (VTE) displayed by one of the most frequently used ventilators that measures exhaled volume at the expiratory valve. Design: Prospective study. Setting: The intensive care units of a pediatric tertiary referral center in London, UK. Patients: A total of 56 intubated children aged between 3 wks and 16.6 yrs who were clinically stable and ventilated with a Servo 300 ventilator. Interventions: The CO2SMO Plus respiratory monitor, which measures flow at the airway opening, was validated using calibrated syringes and appropriate tracheal tubes and connections. Simultaneous in vivo recordings of VTE from the Servo-300 and CO2SMO Plus were compared before (displayed Servo VTE) and after (effective Servo VTepsilon) compensating for ventilator circuit compliance. Measurements and Main Results: The in vitro accuracy of the CO2SMO Plus was within +/-5% over a wide range of volumes and measurement conditions. The displayed Servo 300 VTE overestimated the true VTE by between 2% and 91%. The magnitude of error varied within and between children, according to pressure change (peak inspiratory pressure minus positive end-expiratory pressure), VTE, and circuit size. Mean (SD) error was 32% (20%) in 40 children with displayed Servo VTE of <160 mL and 18% (6%) in 16 subjects with displayed Servo VTE of greater than or equal to160 mL. After correcting for gas compression, effective VTE from the Servo 300 underestimated the true VTE by up to 64% in the smallest infants but continued to overestimate by as much as 29% in older children. Conclusions. The accuracy of tidal volume values is crucially dependent on the site of measurement. Unless measured at the airway opening, displayed values are an inconsistent and misleading indicator of the true volumes delivered.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据