期刊
APPLIED PHYSICS A-MATERIALS SCIENCE & PROCESSING
卷 74, 期 -, 页码 S160-S162出版社
SPRINGER-VERLAG
DOI: 10.1007/s003390101087
关键词
-
The disordered-materials diffractometer D4 at the Institut Laue-Langevin (ILL) has been thoroughly upgraded through improvements in detectors, collimation and shielding to become the D4c instrument. A larger solid angle of detection has increased the total counting rate by a factor of 5, thereby reducing random error in the diffraction measurement, and a corresponding factor of 5 improvement in detector stability has reduced the principal cause of systematic error. The overall precision of the instrument has therefore been increased by a factor of 5 as compared to its previous version, D4b. We present an overview of the D4c instrument's design as well as some results of the successful very high precision experiments performed at D4c since its commissioning in May/June 2000.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据