期刊
ANALYTICAL AND BIOANALYTICAL CHEMISTRY
卷 374, 期 7-8, 页码 1341-1345出版社
SPRINGER HEIDELBERG
DOI: 10.1007/s00216-002-1591-5
关键词
Selenium; Speciation; Preconcentration; Titanium dioxide; Atomic absorption spectrometry
A simple and sensitive method for the selective determination of Se(IV) and Se(VI) in natural water and sludge samples through an adsorptive process on a nanometer-sized TiO2 (anatase) was developed. The conditions for quantitative and reproducible preconcentration, elution, and subsequent GFAAS determination were established. The proposed method gave a concentration factor of 50 for a 100 mL sample volume, characterized by high precision, high reproducibility, and direct determination of Se(IV)/Se(VI). The detection limits (3 sigma, n=11) were 4.7 ng L-1 for Se(IV) and 6.3 ng L-1 for Se (VI); the precision (relative standard deviation) was 0.7% for Se (IV) and 0.9% for Se (VI) at the 0.5 mu g L-1 level.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐