Linearizing an ion cyclotron resonance cell

期刊

REVIEW OF SCIENTIFIC INSTRUMENTS
卷 73, 期 12, 页码 4185-4200

出版社

AMER INST PHYSICS
DOI: 10.1063/1.1518787

关键词

-

向作者/读者索取更多资源

We describe an ion cyclotron resonance (ICR) mass spectrometer that we have built. The design of the instrument was guided in large measure by theoretical consideration; in particular we wished to investigate the effects of improved electrostatic linearity on ICR performance. We found, for instance, that the trap's performance as a mass spectrometer is essentially independent of the trapping potential. By studying both cyclotron and magnetron modes we were able to characterize both the trap and the magnet. Further, we were able to separate effects of image charge from space charge; this gave us measurements of ion number and ion cloud shape. Finally, we show that the instrument readily provides accurate mass measurements with minimal calibration. (C) 2002 American Institute of Physics.