A comparison of a new limited cone beam computed tomography machine for dental use with a multidetector row helical CT machine

出版社

MOSBY-ELSEVIER
DOI: 10.1067/moe.2003.120

关键词

-

向作者/读者索取更多资源

Objective. We sought to compare a new limited cone beam computed tomography (CT) machine for dental use (3DX) with the multidetector CT machine for image quality and skin doses. Study design. Images of the right maxillary central incisor and the left mandibular first molar of an anthropomorphic phantom were taken by both the 3DX and the multidetector CT. A 5-point method was used to evaluate the depiction of cortical and cancellous bone, enamel, dentin, pulp cavity, periodontal ligament space, lamina dura, and overall impressions. Furthermore, the skin doses for both modalities were compared. Results. The image quality of the 3DX was better than the multidetector CT for all items (P < .01). Moreover, the mean skin doses with the multidetector CT were 458 mSv per examination, whereas the doses with the 3DX were 1.19 mSv per examination. Conclusions. These results clearly indicate the superiority of the 3DX in the display of hard tissues in the dental area while substantially decreasing the dose to the patient.