Mechanism of decrease in the atrial potential after implantation of a single-lead VDD pacemaker: Atrial histological changes of a VDD pacemaker lead after implantation in dogs

期刊

PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
卷 26, 期 3, 页码 685-691

出版社

FUTURA PUBL CO
DOI: 10.1046/j.1460-9592.2003.00118.x

关键词

single-lead VDD pacemaker; atrial histological change; insufficient atrial sensing

向作者/读者索取更多资源

HIGASHI, Y., ET AL.: Mechanism of Decrease in the Atrial Potential After Implantation of a Single-Lead VDD Pacemaker: Atrial Histological Changes After Implantation of a VDD Pacemaker Lead in Dogs. The single-lead VDD pacemaker system (VDDPS) enables atria] synchronous ventricular pacing with only one lead in patients with an,atrioventricular block. There are some cases in which the atria] potential decreases after implantation of a VDDPS, making physiological pacing difficult. The mechanism of this decrease has not been elucidated yet. To elucidate the possible relationship between the decrease of the atria] potential after implantation of a VDDPS and histopathological changes of the atrium. We implanted a VDDPS from the jugular vein under anesthesia in 10 adult dogs. The tip of the pacing lead was fixed in the right ventricular apex of the heart under fluoroscopic guidance. Then, the lead was ligated and fixed to the jugular vein at a point where a favorable atria] potential was obtained. The end of the lead was passed from the, neck to the back subcutaneously, then pulled outside and fixed there to measure the atrial potential. The atria] potential was measured using a pacing system analyzer under anesthesia on days 3 (n = 9) and 7 (n = 8), as well as on weeks 2 (n = 6), 3 (n = 4), and 4 (n = 3), after the implantation. The heart was removed from the dogs on day 3 (n = 2), day 7 (n = 2), week 2 (n = 2), and week 4 (n = 4) to examine the atrial histological findings. The atria] potential was 2.7 +/- 0.7 mV at the time of the implantation, 1.7 +/- 1.1 mV (P < 0.05) on day 3, and 1.7 +/- 0.7 mV on week 4 after the implantation. Macroscopically, the pacemaker lead was covered with thrombus, and adhered to the atrial wall in 80% of animals. Microscopically, the endocardium was hypertrophic due to fibrous tissue; besides BBC extravasation, inflammatory cells infiltration and degeneration of myocardial cells, were observed under the endocardium. Inflammatory changes developed in the atrial wall after implantation of the VDDPS, and this seemed to be one of the mechanisms for the decrease of the atrial potential of the VDDPS.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据