4.5 Article

New compact classical 40 kV Mott polarimeter

期刊

REVIEW OF SCIENTIFIC INSTRUMENTS
卷 74, 期 3, 页码 1278-1281

出版社

AMER INST PHYSICS
DOI: 10.1063/1.1535736

关键词

-

向作者/读者索取更多资源

A compact classical electron spin detector based on Mott scattering is-described. This Mott polarimeter has an efficiency of approximate to5.6 x 10(-4), a maximum counting rate of 500 kcps and bulk size 1 cm X 25 cm. The. design of the polarimeter goes back to the classical Mott detector, operating from 100 to 120 kV but it can be combined with conventional analyzers due to its compactness. In this Mott polarimeter an electrostatic acceleration voltage up to 40 kV can be applied and the detectors are energy sensitive silicon diodes operated in reverse bias with variable discriminator threshold. The detectors with the amplifiers are floated on the top of acceleration voltage to allow a field-free travel of the electrons from the scattering gold foil to the detectors., Such features reduce the polarimeters sensitivity to slight motion or changes in the shape of the incoming beam. (C) 2003 American Institute of Physics.