4.8 Article

Retrieval study of uncemented metal-metal hip prostheses revised for early loosening

期刊

BIOMATERIALS
卷 24, 期 6, 页码 1081-1091

出版社

ELSEVIER SCI LTD
DOI: 10.1016/S0142-9612(02)00410-6

关键词

metal-metal total hip replacement; tribologic performance; wear rates; aseptic loosening; third-body wear; SEM

向作者/读者索取更多资源

A tribologic assessment was performed on 22 metal-metal hip prostheses from a single manufacturer, following removal for early aseptic loosening after a mean service life of 32 months (range, 12-59 months). The mean linear wear rate was 7.6 mum/year (range, 2.9-12.8 mum/year). This was below the rates previously observed in other modern metal-metal combinations. A novel contour analysis technique using a coordinate measuring machine showed the mean volumetric wear rate to be 2.02 mm(3) /year (range, 0.55-3.74mm(3) /year), which corresponds to a mean gravimetric wear rate of 16.9mg/year (range, 4.6-31.4mg/year). The mean clearance of 39.8 pm (range, 30-50 mum) was within the optimal range for hard-hard bearing combinations. Evidence of abrasive, adhesive, and third-body wear was found on all bearing surfaces. The tribologic assessment did not indicate manufacturing defects as a cause of early loosening. Equally, third-body wear was too low to be considered a causative factor for early loosening. (C) 2002 Elsevier Science Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据