A technique has been developed to quantify ultratrace Pa-231 (50-2000 ag; 1 ag = 10(-18) g) concentrations in seawater using isotope-dilution thermal ionization mass spectrometry (TIMS). The method is a modification of a process developed by Pickett et al. (Pickett, D. A.; Murrell, M. T.; Williams, R. W. Anal. Chem. 1994, 66, 1044-1049) and extends the technique to very low levels of protactinium. The procedural blank is 16 +/- 15 ag (2sigma), and the ionization efficiency (ions generated/atom loaded) approaches 0.5%. Measurement time is <1 h. The amount of Pa-231 needed to produce Pa-231 data with an uncertainty of +/-4-12% is 100-1000 ag (similar to3 x 10(5) to 3 x 10(6) atoms). Replicate measurements made on known standards and seawater samples demonstrate that the analytical precision approximates that expected from counting statistics and that, based on detection limits of 38 and 49 ag, protactinium can be detected in a minimum sample size of surface seawater of similar to2 L for suspended particulate matter and <0.1 L for filtered (<0.4 mum) seawater, respectively. The concentration of Pa-231 (tens of attograms per liter) can be determined with an uncertainty of +/-5-10% (2sigma) for suspended particulate matter filtered from 5 to 10 L of seawater. For the dissolved fraction, 0.5-1 L of seawater yields Pa-231 measurements with a precision of 1-10%. Sample size requirements are orders of magnitude less than traditional decay-counting techniques and significantly less than previously reported ICP-MS techniques. Our technique can also be applied to other environmental samples, including cave waters, rivers, and igneous rocks.