Thresholds for detecting and characterizing focal lesions using steady-state MR elastography

期刊

MEDICAL PHYSICS
卷 30, 期 4, 页码 495-504

出版社

AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS
DOI: 10.1118/1.1556607

关键词

breast cancer detection; contrast-detail analysis; elastography; elasticity imaging; magnetic resonance imaging; performance characterization

资金

  1. NCI NIH HHS [P01-CA80139] Funding Source: Medline

向作者/读者索取更多资源

An objective contrast-detail analysis was performed in this study to assess the low contrast detectability of a clinical prototype harmonic magnetic resonance elastographic imaging system. Elastographic imaging was performed on gelatin phantoms containing spherical inclusions of varying size and modulus contrast. The results demonstrate that lesions as small as 5 mm can be detected with a minimum modulus contrast of 14 dB. However, the shear modulus of such small lesions was not accurately recovered. In general, the shear modulus of larger focal lesions was accurately (i.e., within 25% of the true value) recovered. The minimum modulus contrast needed to detect focal lesions was observed to decrease with increasing lesion size. (C) 2003 American Association of Physicists in Medicine.