4.4 Article

Radial Force: An Underestimated Parameter in Oversizing Transcatheter Aortic Valve Replacement Prostheses: In Vitro Analysis with Five Commercialized Valves

期刊

ASAIO JOURNAL
卷 64, 期 4, 页码 536-543

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/MAT.0000000000000659

关键词

transcatheter aortic valve replacement; radial force; oversizing; balloon-expandable; self-expandable

向作者/读者索取更多资源

The goal is to inform in depth on transcatheter aortic valve replacement (TAVR) prosthesis mechanical behavior, depending on frame type, design, and size, and how it crucially impacts the oversizing issue in clinical use, and ultimately the procedure outcome. Transcatheter aortic valve replacement is an established therapy for high-risk patients suffering from aortic stenosis, and the indication for TAVR is progressively expanding to intermediate-risk patients. Choosing the optimal oversizing degree is crucial to safely anchor the TAVR valvewhich involves limiting the risks for embolism, aortic regurgitation, conductance disturbance, or annulus ruptureand to increase the valve prosthesis performance. The radial force (RF) profiles of five TAVR prostheses were measured in vitro: the CoreValve 23 and 26 (Medtronic, Minneapolis, MN), the Acurate neo S (Symetis, Ecublens, Vaud, Switzerland), and the SAPIEN XT 23 and 26 (Edwards Lifesciences, Irvine, CA). Measurements were run with the RX Machine equipment (Machine Solutions Inc., Flagstaff, AZ), which is used in ISO standard tests for intravascular stents. Test protocols were adapted for TAVR prostheses. With the prostheses RF profiles' results, mechanical behavior differences could be described and discussed in terms of oversizing strategy and clinical impact for all five valves. Besides, crossing the prostheses' RF profiles with their recommended size windows made the assessment of borderline size cases possible and helped analyze the risks when accurate measurement of patient aortic annulus proves difficult. The prostheses' RF profiles bring new support in clinical decision-making for valve type and size in patients.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据