Impact of metallic dental implants on CT-based attenuation correction in a combined PET/CT scanner

期刊

EUROPEAN RADIOLOGY
卷 13, 期 4, 页码 724-728

出版社

SPRINGER-VERLAG
DOI: 10.1007/s00330-002-1564-2

关键词

PET; CT; attenuation correction; artifacts

向作者/读者索取更多资源

Our objective was to study the effect of metal-induced artifacts on the accuracy of the CT-based anatomic map as a prerequisite for attenuation correction of the positron emission tomography (PET) emission data. Twenty-seven oncology patients with dental metalwork were enrolled in the present study. Data acquisition was performed on a PET/CT in-line system (Discovery LS, GE Medical Systems, Milwaukee, Wis.). Attenuation correction of emission data was done twice, using an 80-mA CT scan (PETCT80) and a Ge-68 transmission scan (PET68Ge)Average count in kBq/cc was measured in regions with and without artifacts and compared for PETCT80 and PET68Ge. Data analysis of region of interests (ROIs) revealed that the ratio (ROIs PETCT80/ROIs PET68Ge) and the difference (ROIs PETCT80 minus ROIs PET68Ge) had a higher mean of values in regions with artifacts than in regions without artifacts (1.2 +/- 0.17 vs 1.06 +/- 0.06 and 0.68 +/- 0.67 vs 0.15 +/- 0.17 kBq/cc, respectively). For most of the studied artifactual ROIs, the PETCT80 values were higher than those of the PET68Ge. Attenuation correction of PET emission data using an artifactual CT map yields false values in regions nearby artifacts caused by dental metalwork. This may falsely estimate PET quantitative studies and may disturb the visual interpretation of PET scan.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据