期刊
JOURNAL OF ANALYTICAL CHEMISTRY
卷 70, 期 11, 页码 1325-1329出版社
PLEIADES PUBLISHING INC
DOI: 10.1134/S1061934815110106
关键词
magnetic molecularly imprinted polymer nanoparticles; copper determination; trace analysis
This paper describes a pre-concentration procedure using magnetic molecularly imprinted polymer (MMIP) nanoparticles for the extraction of copper(II) from aqueous samples prior to its atomic absorption determination. Cu-morin complex was used as template molecule which was chemically bonded to MMIP nanoparticles. MMIP nanoparticles were characterized by scanning electron microscopy as well as Fourier transform infrared spectroscopy. The extraction conditions including pH value, adsorption and desorption time, temperature, amount of ligand and ionic strength of the sample matrix were studied and optimized. The method showed good extraction recoveries (a parts per thousand yen96%) with relative standard deviation below 3.4% for real samples. The linear range of calibration curve was between 5-1000 mu g/L with a correlation coefficient (r (2)) of 0.996. The detection limit was 0.5 mu g/L. The method was suitable for the determination of trace copper in drinking water samples.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据