4.4 Article

Proximal humeral fractures: how stiff should an implant be? A comparative mechanical study with new implants in human specimens

期刊

ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY
卷 123, 期 2-3, 页码 74-81

出版社

SPRINGER
DOI: 10.1007/s00402-002-0465-9

关键词

biomechanical testing; fractures; internal fixation; osteoporostis; proximal humerus

向作者/读者索取更多资源

Background: The objective of this study was to determine the in vitro characteristics of the clinically used and newly developed implants for the stabilization of proximal humeral fractures under static and cyclic loading. The goal was to optimize implant stiffness for fracture stabilization even in weak bone stock. Methods: In a laboratory study using 35 fresh human humeri, the specimens were randomized into 5 groups, which included the clinically used humerus T-plate (HTP), the cross-screw osteosynthesis (CSO), the unreamed proximal humerus nail with spiral blade (UHN), the recently developed Synclaw Proximal Humerus Nail (Synclaw PHN) and the angle-stable Locking Compression Plate Proximal Humerus (LCP-PH). The implant stiffness was determined for three clinically relevant load cases: axial compression, torsion and varus bending. In addition, a cyclic varus-bending test was performed to determine the implant properties under cyclic loading. Results: In contrast to a rather elastic and minimally invasive implant(LCP-PH), the conventionally designed ones (Synclaw PHN, CSO, HTP, UHN) showed rather high stiffness values under static loading. In cyclic loading, a strong decrease in stiffness (p<0.05) was found for the rigid implants HTP and UHN. In comparison with the other implants, only the elastic implant (LCP-PH) showed a significantly lower load reduction in a weak bone stock (17+/-6.2%). Conclusion: The high initial stiffness of rigid implants led to an early loosening and failure of the implant-bone interface under cyclic loading. Implants with low stiffness and elastic characteristics, however, appear to minimize the peak stresses at the bone-implant interface, making them particularly suitable for fracture fixation in osteoporotic bone.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据