期刊
ELECTROPHORESIS
卷 24, 期 7-8, 页码 1271-1275出版社
WILEY-V C H VERLAG GMBH
DOI: 10.1002/elps.200390163
关键词
immobilized pH gradient; isoelectric focusing; two-dimensional polyacrylamide gel electrophoresis
The number of protein spots detected on two-dimensional polyacrylamide gel electrophoresis (2-D PAGE) gels increases as the gel size increases. The largest commercially available systems resolve a few thousand spots, being only a fraction of the total proteome. We have developed an extremely long isoelectric focusing (IEF) system aimed at more complete protein profiling. The system is especially well suited to sensitive detection methods, such as radioactive detection. The major constraint preventing progress in this area has been the inability to create an even density gradient during the immobilized pH gradient (IPG) casting process. We demonstrate for the first time that this constraint can be effectively overcome, to enable greatly increased IEF separating power with all the advantages of IPG technology,