The Effect of Electrode Placement and Interphase Interval on Force Production During Stimulation of the Dorsiflexor Muscles

期刊

ARTIFICIAL ORGANS
卷 38, 期 11, 页码 E142-E146

出版社

WILEY-BLACKWELL
DOI: 10.1111/aor.12346

关键词

Functional electrical stimulation; Muscle force; Peroneal nerve; Dorsiflexors; Foot-drop

向作者/读者索取更多资源

The aims of this study were to investigate whether introducing an interphase interval (IPI) to biphasic pulses during stimulation of the dorsiflexor muscles would affect force production and to determine whether the IPI effect is dependent on electrode position. Twelve healthy volunteers participated in the study. Each subject participated in one session during which electrically induced contraction (EIC) forces of the ankle dorsiflexors were measured with five different IPI settings ranging from 0 to 400s. Forces of EICs were assessed with the electrodes placed either with the proximal electrode positioned over the common peroneal nerve and the second electrode over the dorsiflexor muscles or with both electrodes located over the dorsiflexor muscles. The order of electrode placements and of the different IPI settings was randomized across subjects. The results indicated that the introduction of a 100-s-long IPI may enhance force production when one electrode is located over the common peroneal nerve. However, increasing the duration of the IPI beyond 100s did not result in further increase in force production. In contrast, the introduction of an IPI did not increase force production when both electrodes were located over the dorsiflexor muscles. These findings may help to optimize stimulation settings during functional electrical stimulation to prevent foot-drop.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据