期刊
ANALYTICAL SCIENCES
卷 19, 期 4, 页码 589-592出版社
JAPAN SOC ANALYTICAL CHEMISTRY
DOI: 10.2116/analsci.19.589
关键词
-
A rapid and sensitive method has been proposed for the sequential determination of chromium(III) and total chromium in water samples by flame atomic absorption spectrometry combined with a flow injection on-line preconcentration on a double-microcolumn. The chromium(III) and total chromium in samples were retained on a double-microcolumn with a cation exchange resin, respectively, and eluted directly into a nebulizer by 3 mol L-1 HNO3. The characteristic concentration (gives a derivative absorbance of 0.0044) and the detection limit (3sigma) for chromium were 0.512 mug L-1 and 0.647 mug L-1 for a preconcentration time of 1 min, respectively. This is an improvement of 20 and 14-times than those of conventional FI-FAAS. The proposed method allows the determination of chromium in the range of 0-90 mug L-1 with a relative standard deviation of 3.63% at the 10 mug L-1 level. The method has been applied for the analysis of chromium in reference water of National Research Center for Certified Reference Materials (GBW08607) and other water samples with satisfactory results.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐