4.0 Article Proceedings Paper

Characterization of a polycapillary lens for use in micro-XANES experiments

期刊

X-RAY SPECTROMETRY
卷 32, 期 3, 页码 215-222

出版社

WILEY
DOI: 10.1002/xrs.635

关键词

-

向作者/读者索取更多资源

Polycapillary lenses permit x-rays to be focused down to relatively small (< 50 mum) beam diameters. A polycapillary device, suitable for focusing of synchrotron radiation, was characterized at beam line L of HASYLAB (Hamburg, Germany). Beam parameters such as beam size, transmission efficiency and gain in flux density were determined as a function of the energy-of the primary monochromatic beam entering the polycapillary. At a distance of 4 mm between the lens tip and the sample, the smallest beam diameters were recorded: 30 mum at 7 keV, decreasing to 8 mum at 21 keV. In the energy range from 7 to 15 keV, the transmission efficiency varied between 30 and 45%, while the maximum gain in flux density was calculated to be around 3500. When the energy was changed from 7 to 30 keV, a vertical translation of only 7-8 mum was observed, indicating that this optic is highly suitable for use in mu-XANES investigations. To illustrate this point, the investigation of paper corrosion due to iron gall inks and the speciation of Ni in environmental particles is described. Copyright (C) 2003 John Wiley Sons, Ltd.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.0
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据