4.7 Article

Carbon paste electrode modified with copper(II) phosphate immobilized in a polyester resin for voltammetric determination of L-ascorbic acid in pharmaceutical formulations

期刊

ANALYTICAL AND BIOANALYTICAL CHEMISTRY
卷 376, 期 2, 页码 214-219

出版社

SPRINGER HEIDELBERG
DOI: 10.1007/s00216-003-1866-5

关键词

modified carbon paste electrode; copper(II) phosphate immobilized; voltammetric determination; L-ascorbic acid

向作者/读者索取更多资源

carbon paste electrode modified with copper(II) phosphate immobilized in a polyester resin (CuP-Poly) is proposed for voltammetric determination of L-ascorbic acid in pharmaceutical formulations. The modified electrode allows the detection Of L-ascorbic acid at lower anodic potentials than observed at unmodified electrodes. Several parameters that can influence the voltammetric response of the proposed electrode such as carbon paste composition, pH, scan rate, and possible interference were investigated. The peak current was proportional to the concentration of ascorbic acid in the range 2.0x10(-5) to 3.2x10(-3) mol L-1 with a detection limit of 1.0x10(-5) mol L-1. The stability and repeatability of the electrode for the determination Of L-ascorbic acid are also discussed. Amperometric response was also recorded for electrocatalytic oxidation of the L-ascorbic acid. Concentrations of the vitamin C in pharmaceutical formulations (tablets) measured using the modified electrode and a titrimetric method are in agreement at the 95% confidence level and within an acceptable range of error.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据