4.7 Article

Coronary artery calcium: Accuracy and reproducibility of measurements with multi-detector row CT - Assessment of effects of different thresholds and quantification methods

期刊

RADIOLOGY
卷 227, 期 3, 页码 795-801

出版社

RADIOLOGICAL SOC NORTH AMERICA
DOI: 10.1148/radiol.2273020369

关键词

computed tomography (CT), experimental studies; coronary vessels, calcification; coronary vessels, CT; phantoms

向作者/读者索取更多资源

PURPOSE: To evaluate the effects of different thresholds and quantification methods on the accuracy and reproducibility of coronary calcium measurements with multi-detector row computed tomography (CT). MATERIALS AND METHODS: A cardiac CT phantom containing predetermined calcified cylinders was scanned. Calcium volume and mass were measured at various threshold values ranging from 80 to 230 HU. In 32 patients, two consecutive CT scans were obtained, and the coronary artery calcium score, volume, and mass were measured by one observer at 130- and 90-HU thresholds. Correlation analysis and analysis of variance were performed to evaluate the measurement errors in the phantom study and the interscan variability in the clinical study. RESULTS: In the phantom, mass measurement error varied with threshold and calcium density (P < .01). Mass error was strongly correlated with volume error (r = 0.91, P < .01) but with a much smaller range. In the clinical study, interscan variability of mass measurements was significantly lower than that with other measurement methods for both patients and individual vessels. For the patients, the mean interscan variability of calcium score, volume, and mass at the 130-HU threshold was 20.4%, 13.9%, and 9.3%, respectively. For all methods, interscan variability was not significantly different between the 130- and 90-HU thresholds (P > .05). CONCLUSION: The mass measurement is more accurate, less variable, and more reproducible in coronary calcium quantification than are measurements with other algorithms. Accurate quantification of calcium in each calcified plaque may require that the threshold be set individually, depending on the calcium density. (C) RSNA, 2003.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据