4.7 Article Proceedings Paper

Silver ion-selective electrodes employing Schiff base p-tert-butyl calix[4]arene derivatives as neutral carriers

期刊

SENSORS AND ACTUATORS B-CHEMICAL
卷 91, 期 1-3, 页码 26-31

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/S0925-4005(03)00062-5

关键词

silver ion; sensor; calix[4]arene; Schiff base

向作者/读者索取更多资源

The potentiometric response characteristics and selectivity properties of poly(vinylchloride) (PVC) matrix membrane ion-selective electrodes for silver(I) ion, based on Schiff base p-tert-butyl calix[4]arene derivatives containing N and O as binding sites, were examined. These derivatives proves to work well with silver ions and corresponding electrodes display an excellent response to this ion. These electrodes works well over wide range of concentration (1.0 x 10(-5) to 1.0 x 10(-1) M) with a Nernstian slope. The electrodes have a fast response time of 20 s and operates in the wide pH range. These calix[4]arene derivatives, each showed excellent selectivities in ISE membrane electrodes with log K-Ag,M(pot) around -4.5 against some of the alkali, alkaline and transition metals, even against the strongest interferent, Hg2+ ions, the value was around -3.0. These silver(I) ion-selective electrodes were used as an indicator electrode for potentiometric titration of silver(l) using standard solution of sodium chloride, a sharp potential change occur at the end point. These ISEs were also used to determine low concentration of silver(I) ion in partially non-aqueous medium. (C) 2003 Elsevier Science B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据