4.7 Article

Do airport metal detectors interfere with implantable pacemakers or cardioverter-defibrillators?

期刊

JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY
卷 41, 期 11, 页码 2054-2059

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/S0735-1097(03)00424-8

关键词

-

向作者/读者索取更多资源

OBJECTIVES The aim of this study was to determine whether airport metal detector gates (AMDGs) interfere with pacemakers (PMs) or implantable cardioverter-defibrillators (ICDs). BACKGROUND It is currently unknown whether AMDGs interfere with implanted PMs or ICDs. METHODS A total of 348 consecutive patients (200 PM and 148 ICD recipients) have been tested for the occurrence of electromagnetic interference (EMI) within the electromagnetic field of a worldwide-used airport metal detector. RESULTS No interference, such as pacing or sensing abnormalities, was observed in any of the 200 PM and 148 ICD patients; also no reprogramming occurred. CONCLUSIONS In vivo testing of PM and ICD systems showed no EMI with a standard AMDG. Clinically relevant interactions with implanted PMs or ICDs seem unlikely. (C) 2003 by the American College of Cardiology Foundation.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据