期刊
JOURNAL OF CLINICAL MICROBIOLOGY
卷 41, 期 7, 页码 3167-3174出版社
AMER SOC MICROBIOLOGY
DOI: 10.1128/JCM.41.7.3167-3174.2003
关键词
-
类别
In order to evaluate the LightCycler-based PCR (LC-PCR) as a diagnostic assay technique, a classical pp65 antigenemia assay and the commercially available COBAS Amplicor CMV Monitor (CACM) assay were compared to the LC-PCR assay for the detection and quantitation of cytomegalovirus (CMV) load in 404 parallel specimens of peripheral blood from 66 patients after solid organ transplantation. A good correlation existed among these three assays (r congruent to 0.6, P < 0.0001). The LC-PCR assay was the most sensitive (54% of specimens positive) compared to the CACM (48.6%) and the pp65 antigenemia (26%) assays. The LC-PCR assay detected all samples found positive by using both the CMV pp65 antigenemia assay and the CACM assay. The LC-PCR also had the widest dynamic range (from 250 to 107 DNA copies/ml of plasma). No cross-reactions were found among CMV and Epstein-Barr virus, varicella-zoster virus, or herpes simplex virus in the LC-PCR by using amplification with specifically designed primer pairs. Precision, expressed as the coefficient of variation, was < 3% with standard DNA from cell cultures and between 6.55 and 14.1% with clinical specimens in repeat LC-PCR runs. One run of the LC-PCR took half of the time required for the semiautomated CACM procedure. Because of its sensitivity, specificity, cost-effectiveness, and simplicity, the LC-PCR assay could replace the pp65 antigenemia and the CACM assays as the preferred technique for the surveillance, diagnosis, and monitoring of response of CMV diseases in high-risk populations.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据