期刊
PHYSICS IN MEDICINE AND BIOLOGY
卷 48, 期 13, 页码 2023-2032出版社
IOP PUBLISHING LTD
DOI: 10.1088/0031-9155/48/13/313
关键词
-
Fourier-transform infrared transmission spectroscopy has been used for the determination of glucose concentration in whole blood samples from 28 patients. A 4-vector partial least-squares calibration model, using the spectral range 950-1200 cm(-1), yielded a standard-error-of-prediction of 0.59 mM for an independent test set. For blood samples from a single patient, we found that the glucose concentration was proportional to the difference between the values of the second derivative spectrum at 1082 cm(-1) and 1093 cm(-1). This indicates that spectroscopy at these two specific wavenumbers, alone could be used to determine the glucose concentration in blood plasma samples from a single patient, with a prediction error of 0.95 mM.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐