4.6 Article

Comparative study on conventional and low-flow nebulizers for arsenic speciation by means of microbore liquid chromatography with inductively coupled plasma mass spectrometry

期刊

JOURNAL OF CHROMATOGRAPHY A
卷 1005, 期 1-2, 页码 207-213

出版社

ELSEVIER
DOI: 10.1016/S0021-9673(03)00843-4

关键词

low-flow nebulizers; inductively coupled plasma mass spectrometry; speciation; nebulizers; interfaces, LC-ICP-MS; arsenic

向作者/读者索取更多资源

The performance of conventional and low-flow nebulizer systems with liquid chromatography in differentiating four arsenic species in urine was evaluated. Two low-flow (DIN and MCN) chamber assemblies and a conventional (CFN) nebulizer-spray chamber assembly were compared in the hyphenation of anion-exchange microbore liquid chromatography with inductively coupled plasma mass spectrometry. Under optimal analytical conditions, the detection limits of the four arsenic species were 0.2-0.6 ng ml(-1) for all the nebulizer systems tested. The chromatographic resolution was best in the case of DIN due to its minimal off-column dead volume and superior transport efficiency. Four arsenic species were determined in the certified reference materials NIST SRM 2670E and 2670N. (C) 2003 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足