4.8 Article

Picoamperometric detection of glucose at ultrasmall platinum-based biosensors: Preparation and characterization

期刊

ANALYTICAL CHEMISTRY
卷 75, 期 14, 页码 3308-3315

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac026438u

关键词

-

向作者/读者索取更多资源

A simple method is described for the construction of a glucose biosensor with good reproducibility. After electrochemical etching, the sensing tip of an etched platinum microelectrode was insulated using a synthetic rubber dip coating. The insulating layer was then heat-cured, leading to a small exposed area at the very end of the etched Pt tip, as confirmed by scanning electron microscopy. Phenol and 2-allylphenol were electropolymerized to form an extra insulating layer that effectively retained glucose oxidase (GOX) on the sensing tip of the electrode. On the basis of cyclic voltammetry measurements, the apparent radius of the biosensor tip was estimated to be between 10 and 500 nm, depending on GOX loading. With operational and storage stabilities over 3 weeks, the glucose biosensor prepared using optimal GOX concentration (10 mg/mL) exhibited a picoamperometric current response within similar to2 s and a detection limit of 20 muM with excellent reproducibility.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据