4.8 Article

Performance of pyrolyzed photoresist carbon films in a microchip capillary electrophoresis device with sinusoidal voltammetric detection

期刊

ANALYTICAL CHEMISTRY
卷 75, 期 16, 页码 4265-4271

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac026425g

关键词

-

向作者/读者索取更多资源

Pyrolyzed photoresist films (PPF) are introduced as planar carbon electrodes in a PDMS-quartz hybrid microchip device. The utility of PPF in electroanalytical applications is demonstrated by the separation and detection of various neurotransmitters. PPF is found to form a stable, low-capacitance, durable layer on quartz, which can then be used in conjunction with a microchip capillary electrophoretic device. Sinusoidal voltammetric detection at PPF electrodes is shown to be very sensitive, with a detection limit (S/N=3) of 100 nM for dopamine, corresponding to a mass detection limit (S/N=3) of 2 amol. The selectivity of analysis in the frequency domain is demonstrated by isolating each individual signal in a pair of analytes that are chromatographically unresolved. Effectively decoupling the electrophoresis and electrochemical systems allows the electrodes to be placed just inside the separation channel, which results in efficient separations (80000-100000 plates/m).