期刊
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES
卷 794, 期 1, 页码 157-165出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/S1570-0232(03)00452-5
关键词
protein binding; creatine
Creatine has been quantified in various tissues by a range of methodologies. This paper reports on the development and validation of a simplified HPLC assay to determine plasma creatine, plasma protein binding of creatine, creatine in microdialysate and creatine in over-the-counter products. An isocratic, reversed-phase (C-18) HPLC assay, using potassium phosphate monobasic (pH 4) as a mobile phase, was validated in human plasma and microdialysis perfusion fluid (normal saline). The lower limit of quantification for the assay was 1 mg l(-1) in saline and 5 mg l(-1) in plasma. The RSD was below 6% and accuracy was below 12% in both matrices. Protein binding in human plasma was found to be negligible (<10%). Over-the-counter creatine monohydrate products tested contained 100% creatine monohydrate. This assay was found to be suitable for pharmacokinetic studies and the assessment of plasma creatine and skeletal muscle microdialysate. (C) 2003 Elsevier B.V. All rights reserved.