4.5 Article

Interdisciplinary consensus on the use and technique of vacuum-assisted stereotactic breast biopsy

期刊

EUROPEAN JOURNAL OF RADIOLOGY
卷 47, 期 3, 页码 232-236

出版社

ELSEVIER SCI IRELAND LTD
DOI: 10.1016/S0720-048X(02)00196-1

关键词

breast biopsy; stereotaxy; quality assurance; histopathology

向作者/读者索取更多资源

Purpose: Quality assurance of stereotactic vacuum-assisted breast biopsy (VB). Method: A consensus was achieved based on the existing literature and the experience of VB users (Ethicon Endosurgery, Norderstedt). Results: The imaging work-up must be completed according to existing standards before an indication for stereotactic VB is established. Indications include microcalcifications and small non-palpable masses; for the time being lesions very close to the skin and architectural distortions (radial scar) are considered less suitable. Acquisition of > 20 cores (11 Gauge) should be routinely attempted (goals: as complete a removal of small lesions as possible, thereby increasing diagnostic confidence and reducing so-called 'underestimates'). The pre/post-fire and post-biopsy stereotactic images and a post-biopsy orthogonal mammogram must be documented. All cases with no or uncertain histopathological correlation require discussion in a regular interdisciplinary conference and a documented consensus concerning further work-up or therapy. Standardised documentation of the primary findings and follow-up mammography after approximately 6 months is requested. Conclusion: This consensus includes protocols for the establishment of an indication, performance indicators, interdisciplinary interpretation and therapeutic recommendation, documentation and follow-up. It does not replace official recommendations for percutaneous biopsy. (C) 2002 Elsevier Science Ireland Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据