Marker-based or model-based RSA for evaluation of hip resurfacing arthroplasty? A clinical validation and 5-year follow-up

期刊

ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY
卷 133, 期 11, 页码 1613-1621

出版社

SPRINGER
DOI: 10.1007/s00402-013-1850-2

关键词

Radiostereometric analysis; Marker-based RSA; CAD and RE model-based RSA; Hip resurfacing arthroplasty; Clinical validation; Precision

向作者/读者索取更多资源

The stability of implants is vital to ensure a long-term survival. RSA determines micro-motions of implants as a predictor of early implant failure. RSA can be performed as a marker- or model-based analysis. So far, CAD and RE model-based RSA have not been validated for use in hip resurfacing arthroplasty (HRA). A phantom study determined the precision of marker-based and CAD and RE model-based RSA on a HRA implant. In a clinical study, 19 patients were followed with stereoradiographs until 5 years after surgery. Analysis of double-examination migration results determined the clinical precision of marker-based and CAD model-based RSA, and at the 5-year follow-up, results of the total translation (TT) and the total rotation (TR) for marker- and CAD model-based RSA were compared. The phantom study showed that comparison of the precision (SDdiff) in marker-based RSA analysis was more precise than model-based RSA analysis in TT (p (CAD) < 0.001; p (RE) = 0.04) and TR (p (CAD) = 0.01; p (RE) < 0.001). The clinical precision (double examination in 8 patients) comparing the precision SDdiff was better evaluating the TT using the marker-based RSA analysis (p = 0.002), but showed no difference between the marker- and CAD model-based RSA analysis regarding the TR (p = 0.91). Comparing the mean signed values regarding the TT and the TR at the 5-year follow-up in 13 patients, the TT was lower (p = 0.03) and the TR higher (p = 0.04) in the marker-based RSA compared to CAD model-based RSA. The precision of marker-based RSA was significantly better than model-based RSA. However, problems with occluded markers lead to exclusion of many patients which was not a problem with model-based RSA. HRA were stable at the 5-year follow-up. The detection limit was 0.2 mm TT and 1A degrees TR for marker-based and 0.5 mm TT and 1A degrees TR for CAD model-based RSA for HRA.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据