Detection of awareness in surgical patients with EEG-based indices - bispectral index and patient state index

期刊

BRITISH JOURNAL OF ANAESTHESIA
卷 91, 期 3, 页码 329-335

出版社

OXFORD UNIV PRESS
DOI: 10.1093/bja/aeg188

关键词

anaesthetic techniques, inhalation; anaesthetic techniques, i.v.; monitoring, electroencephalography, BIS; monitoring, electroencephalography, PSI; monitoring, intraoperative

向作者/读者索取更多资源

Background. Patient state index (PSI) and bispectral index (BIS) are values derived from the EEG, which can measure the hypnotic component of anaesthesia. We measured the ability of PSI and BIS to distinguish consciousness from unconsciousness during induction and emergence from anaesthesia and a period of awareness in surgical patients. Methods. Forty unpremedicated patients were randomized to receive: (1) sevoflurane/remifentanil (less than or equal to0.1 mug kg(-1) min(-1)), (2) sevoflurane/remifentanil (greater than or equal to0.2 mug kg(-1) min(-1)), (3) propofol/remifentanil (less than or equal to0.1 mug kg(-1) min(-1)), (4) propofol/remifentanil (greater than or equal to0.2 mug kg(-1) min(-1)). Every 30 s after the start of the remifentanil, patients were asked to squeeze the investigator's hand. Sevoflurane or propofol were given until loss of consciousness (LOC1). Tunstall's isolated forearm technique was used during neuromuscular block with succinylcholine. After tracheal intubation, propofol or sevoflurane were stopped until return of consciousness (ROC1). Propofol or sevoflurane were re-started to induce LOC2. After surgery, drugs were discontinued and recovery (ROC2) was observed. PSI and BIS at LOC (LOC1 and LOC2) were compared with those at ROC (ROC1 and ROC2) (t-test). Prediction probability (P-k) was calculated from values at the last command before and at LOC and ROC. Values are mean (sd). Results. At non-responsiveness, BIS (66 (17)) and PSI (55 (23)) were significantly less than at responsiveness (BIS, 79 (14); PSI, 77 (18); P<0.05). The wide variation with both BIS and PSI measurements of the 80 'awareness' values led to an erroneous classification as unconscious in some cases (BIS, six patients; PSI, nine patients). P-k was 0.68 (0.03) (BIS) and 0.69 (0.03) (PSI). Conclusions. Despite significant differences between mean values at responsiveness and non-responsiveness for BIS and PSI, neither measure may be sufficient to detect awareness in an individual patient, reflected by a P-k less than below 70%.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据