A cryogenic radiometer for absolute neutron rate measurement

期刊

REVIEW OF SCIENTIFIC INSTRUMENTS
卷 74, 期 10, 页码 4280-4293

出版社

AMER INST PHYSICS
DOI: 10.1063/1.1605492

关键词

-

向作者/读者索取更多资源

We detail the design and operation of an electrical substitution radiometer, which measures the absolute fluence of thermal neutron beams based on neutron absorption in a Li-6-rich metallic alloy. We demonstrate that the device possesses a linear response to electrical power in the 50-500 nW range with better than 0.1% absolute accuracy and negligible zero offset. We show the response of the radiometer to a 3 meV neutron beam. We compare the noise of the radiometer to the noise observed upstream in a thin rate monitor and show that the radiometer noise is understood quantitatively. We establish that the radiometer measures neutron rates above 10(5) s(-1) to an absolute accuracy of 0.1%. (C) 2003 American Institute of Physics.