Artefact reduction with alternative cuff configurations

期刊

IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING
卷 50, 期 10, 页码 1160-1166

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TBME.2003.817633

关键词

CAP; cuff-configuration; cuff recordings; EMG; nerve cuff electrode; nerve signals; stimulus artefacts

向作者/读者索取更多资源

In nerve cuff electrode recordings of neural signals, the pick-up of interfering signals can he reduced by choosing appropriate cuff configurations. In the traditionally used tripolar configuration, short circuiting of the end electrodes is expected to reduce the field inside the cuff from interfering signals. A model study suggests that moving the end electrodes toward the center of the cuff reduces the pick-up of interfering signals [9]. In this paper, these properties are studied in more detail using a rabbit model. In addition, a new cuff configuration is suggested, which has an additional set of short circuited end electrodes. The total improvement of signal-to-noise ratio in the new configuration as compared with the traditionally used tripolar configuration was 73% for muscle signals and 127% for the stimulus pulse.