4.4 Article

Improved method for concurrent quantification of antiretrovirals by liquid chromatography-tandem mass spectrometry

期刊

THERAPEUTIC DRUG MONITORING
卷 25, 期 5, 页码 541-543

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/00007691-200310000-00001

关键词

therapeutic drug monitoring; antiretroviral therapy; tandem mass spectrometry; tenofovir

向作者/读者索取更多资源

Therapeutic drug monitoring (TDM) is becoming more widespread to optimize the treatment of patients with HIV/AIDS. The analytic component of TDM requires a drug assay with high specificity, small sample volume requirements, reasonable cost, and rapid turnaround time. This study modified a procedure for the concurrent measurement of 15 antiretrovirals by tandem mass spectrometry. The upper limit of the calibration curves was extended to 10,000 ng/mL, and the matrix for standards was changed from methanol to serum. Also, an additional drug, tenofovir, a nucleotide reverse transcription inhibitor, was included in the revised/improved method. Calibration curves showed good linearity between a concentration range of 100 and 10,000 ng/mL (r > 0.997 for all drugs). Accuracy was assessed by correlation of the calibrators with proficiency testing samples spiked with known drug concentrations and yielded results within 8% of the target values.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据