期刊
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS
卷 67B, 期 1, 页码 655-665出版社
WILEY
DOI: 10.1002/jbm.b.10057
关键词
osteointegration; biomimetic carbonate apatite coating
Biomimetic calcium phosphate (Ca-P) coatings were applied onto dense titanium alloy (Ti6Al4V) and porous tantalum (Ta) cylinders by immersion into simulated body fluid at 37 degreesC and then at 50 degreesC for 24 h. As a result, a homogeneous bone-like carbonated apatitic (BCA) coating, 30 mum thick was deposited on the entire surface of the dense and porous implants. Noncoated and BCA-coated implants were press-fit implanted in the femoral diaphysis of 14 adult female goats. Bone contact was measured after implantation for 6, 12, and 24 weeks, and investigated by histology and backscattered electron microscopy (BSEM). After 6 weeks, bone contact of the BCA-coated Ti6Al4V implants was about 50%. After 12 and 24 weeks, bone contact was lower in comparison with the 6-week implantations at, respectively 24 and 39%. Regarding the BCA-coated porous Ta implants, bone contacts were 17, 30, and 18% after 6, 12, and 24 weeks, respectively. However, bone contact was always found significantly higher for BCA-coated dense Ti6Al4V and porous Ta cylinders than the corresponding noncoated implants. The results of this study show that the BCA coating enhances the bone integration as compared to the noncoated implants. (C) 2003 Wiley Periodicals, Inc.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐