Biocompatibility of a physiological pressure sensor

期刊

BIOSENSORS & BIOELECTRONICS
卷 19, 期 1, 页码 51-58

出版社

ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/S0956-5663(03)00131-3

关键词

biocompatibility; silica glass; fiber optic pressure sensor; intramuscular pressure; tissue response

资金

  1. PHS HHS [R01H03476] Funding Source: Medline

向作者/读者索取更多资源

A newly developed fiber optic micropressure sensor was evaluated for biocompatibility using the International Organization for Standardization (ISO) test standard 10993-6. The test material and an inert control (fused silica glass) were tested in New Zealand white rabbits. Four test specimens were implanted in the paravertebral muscles on one side of the spine about 2-5 cm from the midline and parallel to the spinal column. Similarly, four control specimens were implanted on the opposite side. The implantation periods were 1, 4, and 12 weeks to ensure a steady state biological tissue response. Four animals were tested at each time period. Macroscopic and microscopic observations were performed to compare the biological reactions between the test and control materials. There was an inflammatory reaction at I week which subsided at 4 weeks. There was fibrous tissue growth near the implant that also decreased over time. Most importantly, there was no significant difference in the biological response between the test and control materials. Therefore, we conclude that the pressure microsensor is biocompatible. (C) 2003 Published by Elsevier Science B.V.