4.6 Article

Electric field of a six-needle array electrode used in drug and DNA delivery in vivo:: Analytical versus numerical solution

期刊

IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING
卷 50, 期 11, 页码 1296-1300

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TBME.2003.818467

关键词

electroporation; electric field; finite element; multipole expansion

向作者/读者索取更多资源

We present an analytical solution for the electrical potential and field established by a six-needle array electroporation electrode, which is used in vivo for cancer treatment and DNA delivery. The analytical solution closely matches the,numerical solution obtained with the finite element method: the mean error is less than 0.6%.