4.7 Article

Timing of changes in interstitial and venous blood glucose measured with a continuous subcutaneous glucose sensor

期刊

DIABETES
卷 52, 期 11, 页码 2790-2794

出版社

AMER DIABETES ASSOC
DOI: 10.2337/diabetes.52.11.2790

关键词

-

资金

  1. NCRR NIH HHS [M01 RR00052] Funding Source: Medline
  2. NIDDK NIH HHS [R01 DK 55132] Funding Source: Medline

向作者/读者索取更多资源

The objective of this study was to use a subcutaneous continuous glucose sensor to determine time differences in the dynamics of blood glucose and interstitial glucose. A total of 14 patients with type 1 diabetes each had two sensors (Medtronic/MiniMed CGMS) placed subcutaneously in the abdomen, acquiring data every 5 min. Blood glucose was sampled every 5 min for 8 It, and two liquid meals were given. A smoothing algorithm was applied to the blood glucose and interstitial glucose curves. The first derivatives of the glucose traces defined and quantified the timing of rises, peaks, falls, and nadirs. Altogether, 24 datasets were used for the analysis of time differences between interstitial and blood glucose and between sensors in each patient. Time differences between blood and interstitial glucose ranged from 4 to 10 min, with the interstitial glucose lagging behind blood glucose in 81% of cases (95% CIs 72.5 and 89.5%). The mean (+/-SD) difference between the two sensors in each patient was 6.7 +/- 5.1 min, representing random variation in sensor response. In conclusion, there is a time lag of interstitial glucose behind blood glucose, regardless of whether glycemia is rising or falling, but intersensor variability is considerable in this sensor system. Comparisons of interstitial and blood glucose kinetics must take statistical account of variability between sensors.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据