4.8 Article

Comparison of coronary thermodilution and Doppler velocity for assessing coronary flow reserve

期刊

CIRCULATION
卷 108, 期 18, 页码 2198-2200

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1161/01.CIR.0000099521.31396.9D

关键词

blood flow; coronary disease; microcirculation; pressure

向作者/读者索取更多资源

Background-Thermodilution coronary flow reserve (CFRthermo) is a new technique for invasively measuring coronary flow reserve (CFR) with a coronary pressure wire and is based on the ability of the pressure transducer to also measure temperature changes. Whether CFRthermo correlates well enough with absolute flow-derived CFR (CFRflow) to replace Doppler wire-derived CFR (CFRDoppler) remains unclear. Methods and Results-In an open-chest pig model, CFRthermo was measured in the left anterior descending (LAD) artery and compared with CFRDoppler and CFRflow, measured with an external flow probe placed around the LAD. In 9 pigs, CFR was measured simultaneously by all 3 means in the normal LAD and after creation of an epicardial LAD stenosis. To determine the added effect of microvascular disease, measurements of flow reserve were also performed after disruption of the coronary microcirculation with embolized microspheres. Intracoronary papaverine (20 mg) was used to induce hyperemia. In a total of 61 paired measurements, CFRthermo correlated strongly with the reference standard CFRflow (r=0.85, P<0.001). CFRDoppler correlated less well with CFRflow (r=0.72, P<0.001). Bland-Altman analysis showed a closer agreement between CFRthermo and CFRflow. Conclusion-CFRthermo correlates better with CFRflow than does CFRDoppler.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据