4.8 Article

Biosensor for rapid phosphate monitoring in a sequencing batch reactor (SBR) system

期刊

BIOSENSORS & BIOELECTRONICS
卷 19, 期 3, 页码 233-237

出版社

ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/S0956-5663(03)00209-4

关键词

phosphate biosensor; pyruvate oxidase; thick-film electrode; sequencing batch reactor (SBR); flow-injection-analysis (FIA)

向作者/读者索取更多资源

A thick-film phosphate biosensor based on hydrogel immobilized pyruvate oxidase (POD) has been developed for rapid phosphate process control monitoring in an experimental sequencing batch reactor (SBR) system. We have employed a phosphate biosensor in an off-line monitoring of phosphate concentrations in a bench scale SBR. Measurements with biosensor show a good correlation (r(2) = 0.98) with those of commercial colorimetric phosphate testing kits. The signal response time was 1 min with a detection limit of 5 muM. The biosensor method showed a good operational stability, needed less experimental procedures and a small sample size (similar to20 mul). This allows its practical application for rapid phosphate measurements to obtain real time process data in a SBR system. (C) 2003 Elsevier B.V. All rights reserved.