4.8 Article Proceedings Paper

First human experiments with a novel non-invasive, non-optical continuous glucose monitoring system

期刊

BIOSENSORS & BIOELECTRONICS
卷 19, 期 3, 页码 209-217

出版社

ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/S0956-5663(03)00196-9

关键词

non-invasive blood glucose monitoring; impedance spectroscopy

向作者/读者索取更多资源

This paper describes a non-invasive continuous glucose monitoring system based on impedance spectroscopy. Changes in the glucose concentrations can be monitored by varying the frequency in the radio band over a range, optimised to measure the impact of glucose on the impedance pattern. A number of clinical-experimental studies (hyperglycaemic excursions) were performed with healthy subjects in order to prove the applicability of this approach. The sensor used in these experiments is the size of a wristwatch and holds an open resonant circuit coupled to the skin and a circuit, performing an impedance measurement. In most cases, the experiments showed a good correlation between changes in blood glucose and the sensor recordings. A detailed description of the trials is presented. The results of this first series of experiments can be considered as a proof of concept for this novel non-invasive monitoring approach. Nevertheless, partly due to the indirect measurement, a considerable number of questions remain to be clarified. (C) 2003 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据