Mask ventilation of preterm infants in the delivery room

出版社

BMJ PUBLISHING GROUP
DOI: 10.1136/archdischild-2012-303313

关键词

Neonatology; Resuscitation

向作者/读者索取更多资源

Objective To measure tidal volumes (V-T) and describe the interactions between spontaneous breaths and positive pressure ventilation (PPV) inflations during stabilisation of preterm infants in the delivery room (DR). We used a respiratory function monitor (RFM) to evaluate the first 5min of mask respiratory support provided to preterm infants. Study design An observational study of infants <32weeks gestation, born in a single tertiary perinatal centre receiving mask PPV in the DR. PPV was delivered with a round silicone facemask connected to a T-piece device and RFM. The RFM display was not visible to the resuscitator. Respiratory function parameters in the first 5min after birth were analysed by breath-type (inflations, assisted inflations, spontaneous breaths between PPV, and breaths during continuous positive airway pressure (CPAP)). Parameters measured included V-T, peak inspiratory pressure, peak end expiratory pressure and mask leak. Results A total of 2605 inflations and breaths from 29 subjects were analysed. Substantial leak was observed during all four breath types with median leaks ranging from 24% to 59%. Median tidal volumes were greater during inflations (8.3ml/kg) and assisted inflations (9.3ml/kg) than spontaneous breaths between PPV (3.2ml/kg) and breaths during CPAP (3.3ml/kg). Conclusions Facemask leak is large during resuscitation of preterm infants using round silicone masks. Tidal volumes delivered during PPV inflations are much higher than those generated during spontaneous breathing by an infant on CPAP.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据