4.5 Article

Leak and obstruction with mask ventilation during simulated neonatal resuscitation

出版社

BMJ PUBLISHING GROUP
DOI: 10.1136/adc.2009.182162

关键词

-

向作者/读者索取更多资源

Objectives To evaluate mask technique during simulated neonatal resuscitation and test the effectiveness of training in optimal mask handling. Study design Seventy participants (consultants, registrars and nurses) from neonatal units were asked to administer positive pressure ventilation at a flow of 8 l/min and a frequency of 40-60/min to a modified leak free, term newborn manikin (lung compliance 0.5 ml/cm H2O) using a Neopuff T-piece device. Recordings were made (1) before training, (2) after training in mask handling and (3) 3 weeks later. Leak was calculated. Obstruction (tidal volume <60% of optimal tidal volume) and severe obstruction (<30% of optimal tidal volume) were calculated when leak was minimal. Results For the 70 participants, median (IQR) leak was 71% (32-95%) before training, 10% (5-37%) directly after training and 15% (4-33%) 3 weeks later (p<0.001). When leak was minimal, gas flow obstruction was observed before, directly after training and 3 weeks later in 46%, 42% and 37% of inflations, respectively. Severe obstruction did not occur. Conclusions Mask ventilation during simulated neonatal resuscitation was often hampered by large leaks at the face mask. Moderate airway obstruction occurred frequently when effort was taken to minimise leak. Training in mask ventilation reduced mask leak but should also focus on preventing airway obstruction.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据