Near-infrared reflectance analysis to evaluate the nitrogen and fat content of human milk in neonatal intensive care units

出版社

BMJ PUBLISHING GROUP
DOI: 10.1136/adc.2007.133280

关键词

-

向作者/读者索取更多资源

Objective: To validate near-infrared reflectance analysis (NIRA) as a fast reliable and suitable method for routine, evaluation of human milk's nitrogen and fat content. Setting: One neonatal intensive care unit. Patients: 124 samples of expressed human milk (55 from preterm mothers and 69 from term mothers). Intervention: Measurement of nitrogen and fat content by NIRA and traditional methods (Gerber method for fat and Kjeldahl method for nitrogen). Main outcome measures: Agreement between NIRA and traditional methods. Variability in fat and nitrogen content of human milk. Results: A strong agreement was found between the results of traditional methods and NIRA for both fat and nitrogen content (expressed as g/100 g of milk) in term (mean fat content: NIRA = 2.76; Gerber = 2.76; mean nitrogen content: NIRA = 1.88; Kjeldahl = 1.92) and preterm (mean fat content: NIRA = 3.56; Gerber = 3.52; mean nitrogen content: NIRA = 1.91; Kjeldahl = 1.89) mothers' milk. Nitrogen content of the milk samples, measured by NIRA, ranged from 1.18 g/100 g to 2.71 g/100 g of milk in preterm milk and from 1.48 g/100 g to 2.47 g/100 g in term milk; fat content ranged from 1.27 g/100 g to 6.23 g/100 g of milk in preterm milk and from 1.01 g/100 g to 6.01 g/100 g of milk in term milk. Conclusion: NIRA can be used as a quick and reliable tool for routine monitoring of macronutrient content of human milk and for devising individualised human milk fortification regimens in the feeding of very premature infants.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据