4.4 Letter

A low background neutron flux monitor for the n-TOF facility at CERN

出版社

ELSEVIER
DOI: 10.1016/j.nima.2003.09.060

关键词

neutron flux; silicon detectors; carbon fibre; Monte Carlo simulations

向作者/读者索取更多资源

A small-mass system has been developed for monitoring the flux of neutrons with energy up to 1 MeV at the new time-of-flight facility at CERN, n(-)TOF. The monitor is based on a thin Mylar foil with a Li-6 deposit, placed in the neutron beam, and an array of silicon detectors, placed outside the beam, for detecting the products of the Li-6(n, alpha)H-3 reaction. The small amount of material on the beam ensures a minimal perturbation of the flux and minimizes the background related to scattered neutrons. Moreover, a further reduction of the gamma-ray background has been obtained by constructing the scattering chamber hosting the device in carbon fibre. A detailed description of the flux monitor is here presented, together with the characteristics of the device, in terms of efficiency, resolution and induced background. The use of the monitor in the measurement of neutron capture cross-sections at n(-)TOF is discussed. (C) 2003 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据