Transurethral ultrasound applicators with directional heating patterns for prostate thermal therapy:: In vivo evaluation using magnetic resonance thermometry

期刊

MEDICAL PHYSICS
卷 31, 期 2, 页码 405-413

出版社

WILEY
DOI: 10.1118/1.1639959

关键词

intraluminal; thermal therapy; thermal coagulation; prostate; MR thermal monitoring; ultrasound; hyperthermia

资金

  1. NCI NIH HHS [CA16672-25] Funding Source: Medline
  2. NIDDK NIH HHS [DK51939] Funding Source: Medline

向作者/读者索取更多资源

A catheter-based transurethral ultrasound applicator with angularly directional heating patterns has been designed for prostate thermal therapy and evaluated in canine prostate in vivo using MRI to monitor and assess performance. The ultrasound transducer array (3.5 mm diameter tubular transducers, 180degrees active sectors, similar to7.5 MHz) was integrated to a flexible delivery catheter (4 mm OD), and encapsulated within an expandable balloon (35 mm X 10 mm OD, 80 ml min(-1) ambient water) for coupling and cooling of the prostatic urethra. These devices were used to thermally coagulate targeted portions of the canine prostate (n = 2) while using MR thermal imaging (MRTI) to monitor the therapy. MRI was also used for target definition, positioning of the applicator, and evaluation of target viability post-therapy. MRTI was based upon the complex phase-difference mapping technique using an interleaved gradient echo-planar imaging sequence with lipid suppression. MRTI derived temperature distributions, thermal dose exposures, T1-contrast enhanced MR images, and histology of sectioned prostates were used to define destroyed tissue zones and characterize the three-dimensional heating patterns. The ultrasound applicators produced similar to180degrees directed zones of thermal coagulation within targeted tissue which extended 15-20 mm radially to the outer boundary of the prostate within 15 min. Transducer activation lengths of 17 mm and 24 mm produced contiguous zones of coagulation extending axially similar to18 rum and similar to25 mm from base to apex, respectively. Peak temperatures around 90 degreesC were measured, with similar to50 degreesC - 52 degreesC corresponding to outer boundary t(43) = 240 min at - 15 min treatment time. These devices are MRI compatible, and when coupled with multiplanar MRTI provide a means for selectively controlling the length and sector angle of therapeutic thermal treatment in the prostate. (C) 2004 American Association of Physicists in Medicine.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据