4.5 Article

Active deep brain stimulation during MRI: A feasibility study

期刊

MAGNETIC RESONANCE IN MEDICINE
卷 51, 期 2, 页码 380-388

出版社

JOHN WILEY & SONS INC
DOI: 10.1002/mrm.10699

关键词

deep brain stimulation; functional magnetic resonance imaging; feasibility; safety; implants; RF heating; induced voltage

向作者/读者索取更多资源

The goal of this study was to evaluate the feasibility of active deep brain stimulation (DBS) during the application of standard clinical sequences for functional MRI (fMRI) in phantom measurements. During active DBS, we investigated induced voltage, temperature at the electrode tips and lead, forces on the electrode and lead, consequences of defective leads and loose connections, proper operation of the neurostimulator, and image quality. Sequences for diffusion- and perfusion-weighted imaging, fMRI, and morphologic MRI were used. The DBS electrode and lead were placed in a NaCl solution-filled phantom. The results indicate that there are severe potential hazards for patients. Strong heating, high induced voltage, and even sparking at defects in the connecting cable could be observed. However, it was demonstrated that under certain conditions, safe MR examinations during active DBS are feasible. Certain safety precautions are recommended in this report. (C) 2004 Wiley-Liss, Inc.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据