4D-CT imaging of a volume influenced by respiratory motion on multi-slice CT

期刊

MEDICAL PHYSICS
卷 31, 期 2, 页码 333-340

出版社

AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS
DOI: 10.1118/1.1639993

关键词

4D-CT; motion artifacts; motion characterization; gating and gated imaging

向作者/读者索取更多资源

We propose a new scanning protocol for generating 4D-CT image data sets influenced by respiratory motion. A cine scanning protocol is used during data acquisition, and two registration methods are used to sort images into temporal phases. A volume is imaged in multiple acquisitions of 1 or 2 cm length along the cranial-caudal direction. In each acquisition, the scans are continuously acquired for a time interval greater than or equal to the average respiratory cycle plus the duration of the data for an image reconstruction. The x ray is turned off during CT table translation and the acquisition is repeated until the prescribed volume is completely scanned. The scanning for 20 cm coverage takes about 1 min with an eight-slice CT or 2 mins with a four-slice CT. After data acquisition, the CT data are registered into respiratory phases based on either an internal anatomical match or an external respiratory signal. The internal approach registers the data according to correlation of anatomy in the CT images between two adjacent locations in consecutive respiratory cycles. We have demonstrated the technique with ROIs placed in the region of diaphragm. The external approach registers the image data according to an externally recorded respiratory signal generated by the Real-Time Position Management (RPM) Respiratory Gating System (Varian Medical Systems, Palo Alto, CA). Compared with previously reported prospective or retrospective imaging of the respiratory motion with a single-slice or multi-slice CT, the 4D-CT method proposed here provides (1) a shorter scan time of three to six times faster than the single-slice CT with prospective gating; (2) a shorter scan time of two to four times improvement over a previously reported multi-slice CT implementation, and (3) images over all phases of a breathing cycle. We have applied the scanning and registration methods on phantom, animal and patients, and initial results suggest the applicability of both the scanning and the registration methods. 0 2004 American Association of Physicists in Medicine.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据