期刊
MEASUREMENT SCIENCE AND TECHNOLOGY
卷 15, 期 2, 页码 N21-N24出版社
IOP PUBLISHING LTD
DOI: 10.1088/0957-0233/15/2/N02
关键词
isotope ratio; mass spectrometry; GC/C/IRMS; ratio calculation
A simple algorithm has been derived for the calculation of isotopic composition from gas chromatography/combustion/mass spectrometry. It has the advantage of automatically removing background contributions to the peaks, hence avoiding user subjectivity in placing the baseline. The algorithm has been tested under near ideal conditions of a perfectly resolved peak and also in the more realistic case where there is chromatographic overlap in the peak wings. Performance is shown to be superior to that obtained from the software supplied with the equipment. A spreadsheet implementation of the algorithm has been made available for download via the Internet.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐